Air In Line - Alarm Test; Down Occlusion - Alarm Test - CME BodyGuard 595 Color Vision Operator's Manual

Pain manager infusion system
Table of Contents

Advertisement

Air in Line – Alarm Test
1.
Insert a MicroSet
click is heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
2.
Insert air into the ADMINISTRATION line to allow an air bubble greater than 10
mm to enter into the administration set.
3.
Start infusion as instructed in the "Select protocol" section.
4.
Air in line alarm will be activated as soon as the
air bubble enters the set segment located behind
the pump door.
Press
to turn the alarm off.
Down Occlusion – Alarm Test
1.
Insert a MicroSet
is heard.
2.
Make sure operation temperature is 10-30
NOTE: for instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
3.
Start infusion as instructed in the "Select protocol" section.
4.
Wait 5 minutes and than connect a pressure gauge at a distance of 1 meter from
the exit of the pump. Place both the pump and the pressure gauge at the same
level (on a table).
5.
Wait another 5 minutes and occlude the ADMINISTRATION line downstream of
the pump (between the pump and the pressure gauge).
6.
The Down Occlusion alarm occurs at the time and rates listed below (time is
measured from actual occlusion time to alarm, with accuracy of +30% -20%):
 500 ml/h up to 20 seconds
 60 ml/hr up to 1 minute
 1 ml/hr up to 40 minutes
 0.1 ml/hr up to 11 hours
NOTE: the pressure threshold can be changed to suit patient need (refer to
section on changing the current default Pressure Limit).
BodyGuard 595™Color Vision Operator Manual
TM
administration set and close the door. Press the door until a
TM
into the pump and close the door. Press the door until a click
REGIONAL
Air or Up
Occlusion
Check for air / occlusion
Press
key to mute
o
C
4 – Operation
10:00 -
70%
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents