Automatic Bolus Rate Adjustment; Interpreting The Display Screen During Infusion - CME BodyGuard 595 Color Vision Operator's Manual

Pain manager infusion system
Table of Contents

Advertisement

5.
The display screen will show a graph indicating the
Bolus is being infused. The graphical bar displays
the amount of the bolus to be infused (grey) and
the amount infused (green).
NOTE: once the bolus is completed, the pump will resume the original infusion
program.
NOTE: user can stop bolus at any time by pressing
NOTE: clinician bolus can also be set using 'Clinician Bolus' option on the menu.
It is recommended to select a protocol prior to starting a clinician bolus.

Automatic Bolus Rate Adjustment

The BodyGuard 595 Color Vision™ has a special feature which optimizes bolus rate
during infusion through narrow catheters such as Epidural catheters. The pump will
attempt to deliver the bolus at the pre -set bolus delivery rate. If the pump senses back
pressure above the pre-set occlusion pressure level during bolus delivery it will
automatically reduce the bolus delivery speed until it senses a pressure below the level
at which an occlusion alarm would normally sound. If it still senses an occlusion
pressure above this level at the minimum bolus delivery rate the pump will then alarm
'Down Occlusion'. This feature ensures optimum bolus delivery speed and reduces false
Occlusion alarms during bolus delivery through narrow catheters such as Epidural or
soaker catheters.

Interpreting the Display Screen During Infusion

Protocol &
Protocol Letter
Color coded
Only if air sensor
disable
Basal Rate
(ml/hr or mg/ml)
Time bet ween two patient
boluses (when applicable)
Time bet ween two auto
boluses (when applicable)
BodyGuard 595™Color Vision Operator Manual
REGIONAL
A: FENTANYLCITRATE
air
10
.5
mL/hr
Rate :
Patient Bolus Dose
Lockout Time
Auto-Bolus Dose
Auto-Bolus interval
█████████████████████████████████████
20 of 400 mmHg
EPIDURAL
Bolusing
███████████████████
Bolus Dose
██████████████████████
10:00
- 70%
1.5 mL
20 Minutes
0.5 mL
2 Minutes
4 – Operation
10:00 -
70%
0.363 mL
5 mL
2 mL/hr
20 of 400 mmHg
.
Title bar, color coded
PCA (blue), Epidural
(Yellow), Regional (Green)
Battery Status Indicator,
Actual time, Type of
treatment
Patient bolus volume
(When applicable)
Auto bolus volume
(When applicable)
Actual Pressure Bar
52