2 – Symbols, Warnings and Cautions
Blown fuses could cause a fire hazard. Replace blown fuses on the Charger only
with fuses of the same type and rating (see fuse values on the Charger PCB).
Boluses will be infused up to the full volume that is requested unless the user
stops the bolus manually by pressing
. This is true even when the volume
infused is less than the specified Hour Limit for Volume (e.g. 4 Hr Limit) at the
beginning of a bolus and goes over the limit during the bolus.
Make sure the pump is attached securely to the Charger, which is connected
snugly to an IV pole.
A kinked or occluded administration line may impair the operation of the pump
and the accuracy of the infusion. Before operation, verify that the administration
line is not kinked or occluded.
TM
The BodyGuard 595 Color Vision™ should be operated only with the MicroSet
TM
administration lines. Use of administration lines other than the MicroSet
may
impair the operation of the pump and the accuracy of an infusion.
Drugs must not be administered to the Epidural space unless the drugs are
indicated for this purpose and are administered in accordance with the
indications included in the manufacture's package. Epidural administration of
drugs other than those indicated for Epidural use could result in serious injury to
TM
the patient. For Epidural administration of drugs use MicroSet
only.
Any adjustments, maintenance, or repair of the uncovered pump may impair the
operation of the BodyGuard 595 Color Vision™ Pain Manager and/or the
accuracy of the infusion. Any adjustments, maintenance, or repair of the
uncovered pump or charger should be performed by authorized skilled
technicians. Any adjustments, maintenance, or repair of the uncovered pump or
charger while connected to the power should be avoided.
The BodyGuard 595 Color Vision™ Pain Manager should be operated within a
temperature range of 10C (50F) to 45C (113F) and up to 85% humidity.
Operating the pump at temperatures and/or humidity other than within that range
may affect accuracy.
Unsafe operation may result from using improper accessories. Use only
accessories and options designed for this system.
Disposables must be compatible with the medicine delivered.
Battery charging is enabled as long as the charger cord is connected to the
mains and the pump is in the charger. Switching the pump off does not
disconnect it from the mains. To disconnect from mains, remove the charger
cord from mains. To disconnect pump from mains, remove it from the charger.
Dropping the BodyGuard 595 Color Vision™ Pain Manager could cause damage
to components. If the pump is dropped, return the pump for inspection by
qualified service personnel.
Use aseptic technique. Patient infection may result from the use of non-sterile
components. Maintain sterility of all disposable components and do not re-use
single use administration sets.
BodyGuard 595™Color Vision Operator Manual