6.
The display screen will show a graph indicating the
bolus is being infused. The graphical bar displays
the amount of the bolus to be infused (grey) and
the amount infused (green).
NOTE: once the bolus is completed, the display will change back to the menu.
NOTE: user can stop bolus at any time by pressing
NOTE: when MediGuard
Warning: boluses will be infused up to the full volume that is requested unless
the user stops the bolus manually by pressing
volume infused is less than the specified Hour Limit for Volume (e.g. 4 Hr Limit)
at the beginning of a bolus and goes over the limit during the bolus.
To infuse a Clinician Activated Bolus during an infusion, perform the following steps:
1.
During operation, press
initiate the clinician bolus.
2.
Enter relevant code and press
3.
Enter the volume of the clinician bolus and press
to begin priming procedure. You may stop
the bolus at any time by pressing
4.
The volume for the bolus will be in ml, mg or µg
depending on the settings of the existing protocol
entered for the relevant infusion.
BodyGuard 595™Color Vision Operator Manual
TM
is ON clinician bolus cannot be set from the menu.
and then press
.
.
EPIDURAL
10:00 -
Bolusing
0.363 mL
███████████████████
Bolus Dose
██████████████████████
20 of 400 mmHg
. This is true even when the
to stop the current infusion and
EPIDURAL
10:00 -
Clinician Bolus
Dose
5
mL
(Ma ximum
20mL)
Enter and Press
EPIDURAL
10:00 -
Clinician Bolus
Dose
5
mcg
(Ma ximum
40mcg)
Enter and Press
4 – Operation
█
70%
■
5 mL
2 mL/hr
█
70%
■
Key
█
■
70%
Key
51