Bumper Installation (For Installations With Bassinets - GE PANDA Service Manual

Warmers
Table of Contents

Advertisement

Appendix E: Installing Wall Mount Units

Bumper Installation (for Installations with Bassinets)

E.3 Bumper Installation (for Installations with Bassinets)
The bumper is intended to help the clinician properly position the bassinet during use. Bassinets are
available in many different sizes and configurations. Confirm that the center of the mattress is 19" (48.3 cm)
away from the front of the rails on the warmer with the bumper installed. If the included bumper is not the
right dimension for the facility's bassinets, the facility may use an alternative bumper design to ensure
proper positioning.
1.
Position the bumper on the wall such that it is leveled horizontally and its center is aligned with the
center of wall mount bracket and the bumper holes are 7" (17.8 cm) from the floor.
Note:
For bassinets with a bumper being at a height different from 7" (17.8 cm), the bumper distance from
the floor can be adjusted to match the bassinet bumper height.
2.
Use the bumper as a template to mark the mounting holes on the wall.
3.
Use a ¾" (19mm) diameter metal cutting hole saw or drill four holes through the dry wall and through
metal drywall tracks behind the wall.
4.
Secure the two Toggler® anchor bolts as instructed in step 4 of
5.
Mount the bumper on the wall with the flat washers (6600-0173-400), split ring washers
(6600-0440-400) and screws (660-0414-400) and screws provided (see
with approximately 80 lb-in of torque.
E-6
section
E.2.
Figure
E-4). Tighten the screws
Service Manual