Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
D.1.1.1 Unit Information Displays
•
Unit type
•
Software
•
Processor and option versions
•
PC Board ID's
•
Battery voltage status
•
Power fail minutes
•
Language
D.1.2 Operator Settings
Operator Settings enable users to customize the behavior of alarms, user functions, and displayed units by
setting default values for the parameters and options listed in the following table.
Parameter/Option
Patient temp alarm
CPAD (check patient alarm disabled)
Man temp alarm
Touchless silence
Bed elevate (for Bedded Warmers with
Elevating Base)
Bed elevate (for Bedded Warmers with
Fixed-Adjustable Base, Freestanding
Warmers, and Wall Mount Warmers)
Temp Units
Scale Units
Scale Resolution
Time Format
Low pulse limit
High pulse limit
Low sat limit
High sat limit
Masimo SpO
Masimo SpO
Nellcor SpO
Nellcor SpO
Pulse tone volume, Panda unit
Pulse tone volume, Giraffe unit
Alarm volume
Language
D-2
averaging
2
sensitivity
2
Response Mode
2
Display
2
Settings
0.5C / 1.0C
5% - 50%
Off / 0.5C / 1.0C
Enable / Disable
Enable / Disable /
Disengaged
Enable / Disable /
Disengaged
C / F
grams / lbs-oz
10/5 grams
12/24
0 - 110
111 - 240
0 - 99
86 - 105
2-4 / 8 / 16
Normal / Maximum
Normal / Fast
Pleth / Blip
0/ 1 / 2 / 3 / 4
0/ 1 / 2 / 3 / 4
1 / 2 / 3 / 4 Default is 4.
Many
Default
1.0C
25%
Off
Enable
Enable
Disengaged
C
grams
10 grams
24
100
200
85
100
8
Normal
Normal
Pleth
1
0
3
English
Service Manual