1.3.14 Scale (Bedded Models Only)

The scale communicates with the main processor via RS-485. (Refer to
scale cable is in the probe panel housing. The scale is powered by 12V DC supplied by the control board.
The power circuit includes over-current protection.
1.3.15 SpO
2
The Pulse Oximetry (SpO
For units that have the SpO
in the probe panel housing. (Refer to
an SpO
interface board. Oximetry parameters and alarms are passed to the main application processor
2
via serial communication. The OEM module output is communicated to the user visually or audibly as
required.
Service Manual
FIGURE 1-16. Scale Interface
) option utilizes an OEM module commonly used in other monitoring equipment.
2
option, the SpO
patient probe connects to the SpO
2
2
Figure
1-17.) The probe panel housing contains the OEM module and
FIGURE 1-17. Pulse Oximetry (SpO
Chapter 1: Functional Description
System Functions
Figure
1-16.) The connector for the
module which is located
2
)
2
1-23