Diagnostics - GE PANDA Service Manual

Warmers
Table of Contents

Advertisement

Chapter 4: Troubleshooting
Service Mode

4.1.7 Diagnostics

Diagnostics provides the following functions:
Temperature A/D Channels
VGA Test
Touch Panel Buttons / Switches
Alarm Tests
Odometer
Logs
Heat Engine Check
4.1.7.1 Temperature ADC Diagnostics
This screen displays reference voltages, readings from both patient probe thermistors, one fixed thermistor
input, and the ambient temperature of the Control Board.
Expected Values
ADC Channel
1
Vthref
2
Vo/t
3
Pat Probe Thermistor 1
4
Pat Probe Thermistor 2
5
Vref 37°C
6
Ambient Temp
7
Unused
8
Vref 1P235
4-6
Raw Value
Corrected Value
2048 (+ / - 4)
1500 (+ / - 3) mV
NA
NA
NA
Pat Probe Thermistor 2 ±0.3°C
NA
Pat Probe Thermistor 1 ±0.3°C
2232 (+ / - 23)
37.3 (+ / - 0.2)°C
NA
NA
NA
NA
1706 (+ / - 8)
1250 (+ / - 6) mV