1.
Before installing the heater head, confirm that the unit serial number on the rear side of the head
matches with the serial number on the wall bracket.
Warning: Heater alignment may be lost if warmer head is removed and reinstalled
on a different wall bracket. Always confirm serial number matching between the
heater head and the wall bracket before re-installation.
2.
For installations to be used with bassinets, confirm the wall mount bracket is positioned correctly by
verifying that the top edge of the bracket is 60.8" +/- 0.1" (154.4 cm +/- 0.25 cm) from the floor. For
installations to be used with fixed surfaces, confirm the wall mount bracket is positioned correctly by
verifying that the top edge of the bracket is 24.8" +/- 0.1" (63 cm +/- 0.25 cm) from the floor (See
Figure
E-2).
3.
The steps below are required to perform heater alignment which is necessary upon installation of wall
mount models:
Warning: The heater head is 15 kg (33 lbs). Follow your local health and safety
guidelines fro manual handling of heavy objects when lifting and carrying the
heater head.
Note:
Heater alignment is not required if heater head is re-installed on its corresponding wall bracket on
which the heater has been aligned before.
a.
Remove the upper heating housing as instructed in section 5.3.1 to access the heater alignment
screws.
b.
Mount the heater head by inserting the shoulder screws on the back of the header head into their
corresponding bracket slots on the wall bracket. Install the four M6 button head screws.
c.
Fit the alignment laser in to the laser holder. Turn laser about 1/8th turn to seat it in the bore of the
laser holder.
d.
Locate and mark the alignment spot as illustrated in Figure 5-3.
For installations to be used with bassinets, the alignment spot shall be at 40.5" +/- 0.1" (103 cm +/- 0.25
cm) distance from the wall and at 0.75" +/- 0.1" (1.9 cm +/- 0.25 cm) distance from the center of the
warmer unit.
For installations to be used with fixed surfaces, the alignment spot shall be at 25.75" +/- 0.1" (65 cm +/-
0.25 cm) distance from the wall in line with the center of the warmer unit.
Note:
Use a laser level (with two wall leveling capability) or a combination of T-square and Plumb-bob to
locate the alignment spot accurately (See
Note:
Make sure the alignment spot location is measured from the wall, not the baseboard.
e.
Use a long 3 mm hex key to adjust the dish's 3 adjustment screws located in the bosses
underneath the cotter pins. Loosening or tightening the 2 screws on either side will move the laser
spot from left to right, with longitudinal shift, and adjusting the screw in back moves the spot up
and down in the target. Adjust the side screws first, then adjust the bottom screw, so that the laser
spot is on the marked alignment spot on the floor.
f.
Remove the four M6 button head screws. Lift the heater head up slightly to release the shoulder
screws from the bracket slots and remove the head from the bracket.
g.
Place the heater head on a flat surface and re-install the upper heating housing.
4.
Mount the heater head by inserting the shoulder screws on back of the head into their corresponding
bracket slots. Secure the bracket to the heater head using the four M6 button head screws provided.