Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
Byte 5 – Manual Temp Alarm Setting.
"SPO2SET"
With Masimo SpO
installed:
2
Byte 0 – The Pulse Tone Volume.
Byte 1 – Averaging (0x0 for "2-4", 0x02 for "8", and 0x06 for "16").
Byte 2 – Sensitivity (0x0 for Maximum, 0x01 for Normal).
Byte 3 – Low Pulse Rate Limit Setting.
Byte 4 – High Pulse Rate Limit Setting.
Byte 5 – Low Saturation Limit Setting.
Byte 6 – High Saturation Limit Setting.
With Nellcor SpO
installed:
2
Byte 0 – The Pulse Tone Volume.
Byte 1 – Response Mode (0x01 for Normal, 0x02 for Fast).
Byte 2 – SatSeconds™ Limit Setting.
Byte 3 – Low Pulse Rate Limit Setting.
Byte 4 – High Pulse Rate Limit Setting.
Byte 5 – Low Saturation Limit Setting.
Byte 6 – High Saturation Limit Setting.
Service Manual
D-7