Page 3
MSA Permanent Instrument Warranty 1. Warranty- Seller warrants that this product will be free from mechanical defect or faulty workmanship for a period of eighteen (18) months from date of shipment or one (1) year from installation, whichever occurs first, provided it is maintained and used in accordance with Seller’s instructions and/or recommendations.
Page 4
5. Do not paint the Chillgard RT System. Also, do not paint near any of the sample line inlets to ensure paint is not deposited on the sample inlet fitting of the units. Such paint deposits would interfere with the sampling process, whereby a sample of the atmosphere being monitored is drawn into the Chillgard RT System.
Page 5
9. Properly vent the exhaust of the Chillgard RT System to a safe area. Improper venting of the exhaust may cause serious personal injury or death. 10. Ensure the gas sample to the system is clean and non-condensing. Install end-of-line filters in all sample inlet lines.
The Internal Multipoint Sequencer Unit: ....... 1-7 Terminology ..............1-8 Applications ..............1-9 Sensor Selectivity* ............1-9 Table 1-2. Typical Chillgard RT System for R-11 Cross-Sensitivity Response Data ... 1-9 Table 1-3. Typical Chillgard RT System for R-12 Cross-Sensitivity Response Data ..1-10 Table 1-4.
Page 7
R-500 Cross-Sensitivity Response Data ..1-12 Table 1-10. Typical Chillgard RT System for R-502 Cross-Sensitivity Response Data ..1-13 Table 1-11. Typical Chillgard RT System for Ammonia* Cross-Sensitivity Response Data ....1-13 Table 1-12. Typical Chillgard RT System for R-404a Cross-Sensitivity Response Data ..1-14 Chapter 2 Installation and Set-up ........
Page 8
Wiring the Chillgard RT Unit ........2-11 Analog Signal Output Wiring (FIGURE 2-5) ....2-11 4-20 mA ..............2-11 0-10 VDC ..............2-11 RS 232 Output ............2-12 Figure 2-5. Wiring Connections ......2-12 Ferrite Bead Installation ..........2-13 Figure 2-6. Ferrite Bead Installation ....... 2-13 CAUTION ............
Page 10
Table B-1. Specifications…………………………….B-3 Figure B-3. Relay Module Main Power Connector and Communication Cable Connector Locations………………………………………...B-3 Communication Cable Connections…………………B-4 Figure B-4. Chillgard RT Relay Module Power Wiring Diagrams for 120 Volts and 240 Volts ………………………………………..B-4 Figure B-5. Chillgard RT Refrigerant Monitor………B-4 viii...
Page 11
Communication Cable Connections ......B-7 WARNING ..............B-7 CAUTION ..............B-7 WARNING ..............B-7 Communication Cable Connections to the Chillgard RT Relay Module ........B-8 Communication Cable Connections to the Chillgard RT Refrigerant Monitor ......B-8 Relay Wiring ..............B-8 Operation ................B-9 Front Panel ..............
Page 12
Multi-refrigerant Capability* ......D-1 Introduction ..............D-1 Specifications ..............D-1 Table D-1*. Multi-Refrigerant Chillgard RT Refrigerant Monitor ...... D-1 Table D-2. Typical Cross Sensitivity at 100 Parts Per Million (PPM) at 25°C for the Chillgard RT Refrigerant Monitor Calibrated at 1000 PPM R11 as Expressed in Response Factors ....
Page 13
Installation ............... D-2 Figure D-1. Label and RS-232 Location ....D-3 Figure D-2. Calibration Label ........D-4 Figure D-3. Manufacture Date Label ......D-4 Initial Setup ............... D-5 Calibration ............... D-5 Display Screens .............. D-5 WARNING ............. D-5 Figure D-4. Gas by Point Setup Screens ....D-7 Figure D-5.
These instruments can provide continuous refrigerant gas monitoring (see TABLE 1-1 for Operating Specifications). All units listed below have been evaluated by Underwriters Laboratory. The ammonia units have not been evaluated by Underwriters Laboratories, Inc. CHILLGARD RT MONITORS EVALUATED BY UNDERWRITERS LABORATORIES, INC. R-11 R-502 R-12 R-123...
Table 1-1. Chillgard RT System General Operating Specifications PERFORMANCE SPECIFICATIONS STABILITY 0-50 ppm, +1 ppm; ammonia is 2 ppm 51-1000 ppm, +10% of reading SHORT TERM NOISE +1 ppm peak over a 10-minute period; ammonia* is +2 ppm LINEARITY 0-50 ppm +1 ppm; ammonia* is +2 ppm...
**For UL2075 testing from 50°C to 61°C, the Temperature Effect is +0.6%/°C of reading and +1 ppm/°C For the Chillgard RT Remote Relay* unit option, see Appendix B, "Relay Module." The Chillgard RT Monitor is intended for indoor use in mechanical equipment rooms housing air conditioning or refrigeration equipment or for use in areas where bulk refrigerants are stored.
If the end of sample line filters are not included, they must be purchased. Place an end-of-sample line filter (P/N 711561). Figure 1-5. Chillgard RT Monitor with Optional Beacon General Description The Chillgard RT Unit: •...
• RS 232 output (see Appendix C for more information). The Internal Multipoint Sequencer Unit: • Allows a single Chillgard RT unit to monitor up to four or eight areas or locations, if equipped • Each point is capable of sampling from remote areas up to 500 feet (152 m) from the Chillgard RT unit •...
PUMP- The electric motor driven device that moves the gas sample to the Chillgard RT Monitor. ALARMS- The Chillgard RT System has three alarms to alert the user at specific, user-adjustable refrigerant gas concentrations. (Level 1 = Caution, Level 2 = Warning and Level 3 = Alarm.) RELATIVE HUMIDITY- The percent of water vapor saturation in air at a given temperature.
Applications The Chillgard RT Monitor is factory-calibrated for a particular refrigerant gas. When calibrated, the Chillgard RT System is used in a wide variety of industrial applications, such as: • detection of refrigerant gases leaking into the atmosphere • ambient air monitoring for health and safety purposes...
Retain the original packing form for re-use in the event the unit must be returned for service. 3. Using a screwdriver, loosen the two latches on the Chillgard RT door (FIGURE 2-1) by turning 1/4 turn counterclockwise. 4. Open the front door.
Figure 2-1. Front Door Latches Location of the Monitor WARNING This unit must not be located in areas that may contain a flammable mixture of gas and air; otherwise, an explosion may occur. The monitor performance is dependent on its location and sensing pick- up location(s).
4. Do not mount unit where it will be exposed to direct solar heating. 5. For proper cooling, allow at least three inches (7.6 cm) clearance around all surfaces except for the mounting surface. 6. Mount the unit so the front panel is easily seen and accessed for service and calibration.
Activity in the Room The expected activities in the room must also be considered when determining the sensing point. Refrigerants will concentrate near the floor. Typically, sampling 12-18 inches (30.5-45.7 cm) above the floor is sufficient for early warning and to provide adequate protection for someone working close to the floor.
8. DO NOT: • back-pressure the exhaust line or • connect it to a vacuum source or • install a flow meter in the exhaust line. ALWAYS : connect an exhaust line that is vented to a safe area or an outside atmosphere.
Wiring Connections Opening the Unit All wiring to the Chillgard RT unit is made via the side entries. Open the unit to provide complete access to all wiring connections. Ensure that all wiring codes are followed. These codes include, but are not limited to, the National Electrical Code.
1. Using a screwdriver, loosen the two latches on the enclosure door (FIGURE 2-1). 2. Open the front door. 3. Determine the power requirements for your Chillgard RT unit, ensuring the power is clean and reliable. (Refer to TABLE 1-1 for current capacity specifications.)
L1 and L2 and ground wire to terminal. Analog Signal Output Wiring (FIGURE 2-5) The Chillgard RT unit has two available analog outputs. Each analog output is software-selectable between 10% or 100% of full scale: • 4-20 mA, isolated, current sourcing •...
The voltage output can be programmed to correspond to the channel presently being sampled. The front-panel screen shows: 1V/PT IF OUT = 1-10V (indicating one volt per point if output equals one to 8 volts). When enabled, and the unit is monitoring from: •...
Ferrite Bead Installation Some models are shipped with ferrite beads. Follow the instructions below for proper installation. • The input and output signals must be connected by using twisted, shielded wire pairs. • The shield or its drain wire (FIGURE 2-6) must be connected to the grounding terminal on the power supply plate.
Optional Alarm Beacon Your unit may have an optional alarm kit installed. This kit is made up of a beacon on the top of the unit. It is factory-wired so no additional wiring is necessary; it lights when any alarm indication is given by the instrument.
Trouble Relay There is one relay (the Trouble relay) within the unit that indicates that a trouble or start-up condition exists. This relay is configured from the factory and operates differently than the alarm relays. It operates in a normally-energized mode. This relay is energized when the instrument is: •...
AWG. Shielded wire is necessary. Ground the shields of all cables at the transmitting end of the signal. Do not ground or connect the shield at the Chillgard RT Refrigerant Monitor. CAUTION All field wiring must be done in accordance with national and local electrical codes.
• the monitor returns to the previous sequencer setup (FIGURE 3-4). Chillgard RT operation with a sequencer has a normal dwell time of 30 seconds per sample line, or four measurement cycles. To improve monitor performance in noisy environments and eliminate false alarms due to high noise, a user-selectable parameter "threshold"...
Chapter 3 Start-up and Initial Adjustments Before applying power to the Chillgard RT Refrigerant Monitor, follow the Appendix A check list. This will help ensure that the instrument is properly configured for operation. All instrument operation is performed via the front panel which consists of four keys and a two-line by 20-character vacuum fluorescent display.
Perform the calibration procedure regularly and maintain a log of calibration adjustments. Calibration frequency may increase for a variety of reasons. If calibration cannot be performed at any step, STOP; consult MSA at 1-800-MSA-INST.
Calibration Equipment Equipment needed: • Calibration Kit (MSA ATO #50; FIGURE 4-1) • A SPAN gas cylinder • Optional ZERO gas cylinder. • A ZERO gas cylinder may not be needed. • For ammonia units, zero gas cylinder must be used.
The Calibration Kit contains a ZERO gas scrubber which can be used in place of a ZERO gas cylinder if the ambient air around the Chillgard RT contains little or no refrigerant. CAUTION The zero gas scrubber must be replaced periodically. The frequency of replacement depends on the concentration of the ambient refrigerant vapors.
Calibration Procedures Before calibrating the Chillgard Monitor, leak-check the sample line(s) connected to the monitor: 1. Temporarily block the sample inlet at the end-of-line filter(s) and verify that the monitor gives a trouble alarm. • If the monitor is a Multipoint Sequencer, the trouble alarm will not activate until the blocked sampling point is selected by the Sequencer.
• If adjustments are needed, press ADJ to access the UP, DOWN and ENTER keys; enter the zeroing and spanning adjustments. 4. Zeroing the Monitor: When zero gas is required, attach a zero gas scrubber or zero gas cylinder to the connector as shown in FIGURE 4-3. NOTE: Use only the zero gas cylinder on monitors calibrated for ammonia.
NOTE: The span gas must be appropriate for monitor calibration. For many applications, a span gas cylinder of the refrigerant of interest is available from MSA. If a span gas cylinder of the specific refrigerant is not available, MSA has determined an appropriate synthetic span gas.
6. Perform the TABLE 4-2 sequence. a. Apply zero gas when requested. • Allow enough time for a stable reading. b. Accept (OK) or adjust ("ADJ") and enter a zero reading. c. Apply span gas when requested. • Allow enough time for a stable reading. d.
Chapter 5 Maintenance General Maintenance Under normal operation conditions the Chillgard RT Monitor requires minimal maintenance. The end-of-sample-line filter(s) should be examined regularly for dirt/dust build-up and replaced when necessary. If the filter(s) become severely clogged with dust, the flow rate will eventually fail and a flow alarm will occur.
Table 5-1. Replacement Parts PART PART Internal Filter, Single Point 711561 Internal Filters, all Multipoint and single-point Ammonia 655533 Filter, End-of-Line 711561 Ammonia Unit, Filter, End-of-Line 655652 2.0 amp, 250 volt Fuse 638091 Display 655786 Front Panel Printed Circuit Board Assembly (grey ribbon cable) 10014705 Power Supply 637394...
2. Check span gas supply. 3. Check all tubing, filters and fittings for leaks. 4. For sequencer units, make sure unit is locked on the selected sample port used for calibration. 5. Leak test the flow system. 6. Return to MSA for service.
Page 86
OFF and re-start unit. sample flow line at 2. Leak test the flow system. unit startup Case temperature Main board 1. Return to MSA for service. failed component Case temperature Case 1. Return to MSA for service. range failed temperature is 0 to 90 °C...
Page 87
1. Replacement requires MSA service. range failed temperature range of 45 to 90 °C Humidity sensor failed Checks H 2 O 1. Return to MSA for service. sensor output Infrared source failed Checks the I 1. Return to MSA for service of source...
Chapter 6 Data Logging Log Size The Chillgard RT unit allows 1,078 sets of data and events to be logged. This data is either the maximum or average of the "last" data calculated. Last Values The maximum value is taken from the ’last’ data; this is not always the highest value displayed, but the last valid data taken from a point.
Page 89
Appendix A Check List Before applying power to the Chillgard RT Refrigerant Monitor, check for all items in the following table: ITEM CHECK FOR: CHECKED OR INITIALED Proper mounting of Chillgard RT Refrigerant Monitor on a non-vibrating surface which does not block the front panel.
This accessory module has not be evaluated by Underwriters Laboratory, Inc. General Description The Chillgard RT Relay Module is intended to provide discrete relay outputs for the Chillgard RT Refrigerant Monitor equipped with a multi- point sequencer. The Chillgard RT Refrigerant Monitor has three alarm threshold levels (CAUTION, WARNING, and ALARM), each with a corresponding form C relay output.
Monitor and sets its installed relays according to information received. While the Chillgard RT Relay Module receives its information from the Chillgard RT Refrigerant Monitor, it is important to note that the Relay Module operates independently with respect to the Reset Switch and Audio operation.
Specifications Table B-1. Specifications TEMPERATURE OPERATING 0 to 50°C RANGE STORAGE -20 to 60°C POWER 85 - 264 Volts @ 40 Watts max. MAXIMUM 30 seconds WARM-UP TIME HUMIDITY 0 to 99% RH non-condensing SYSTEM COMMUNICATION RS-485 RELAYS 10 Amps @ 125, 250 VAC COMMON TROUBLE 10 Amps @ 30 VDC AND HORN RELAY...
Page 95
Figure B-4. Chillgard RT Relay Module Power Wiring Diagrams for 120 Volts and 240 Volts AC Figure B-5. Chillgard RT Refrigerant Monitor Communication Cable Connections...
Identifying the Unit Each bank of relays (see FIGURE B-3) is assigned to a particular alarm level. Chillgard RT Relay Modules can be purchased with one, two, or three internal relay banks (TABLE B-2). Table B-2. Part Numbers and Supplied Relays...
Identify the main power voltage that will power the Chillgard RT Relay Module; if unsure of your primary power, consult an electrician or local power utility. WARNING Application of improper power to the Chillgard RT Relay Module will result in damage to the unit.
Relay Wiring Before installing any wiring to the relays within the Chillgard RT Relay Module, refer to the manual for the specific equipment being connected to the relays for wiring procedures and requirements. If your unit is a four-point Chillgard RT model, use the first four relays.
Section 4, Troubleshooting Guidelines. Trouble Indicator The Trouble Indicator turns ON and remains ON when a trouble condition is detected. The trouble condition may exist at the controlling Chillgard RT Refrigerant Monitor or it may exist locally at the Chillgard RT Relay Module.
• This indicator remains OFF during the initial Chillgard RT Refrigerant Monitor warm-up, setup and calibration. Chillgard RT Relay Module Relays There are three types of relays in the Chillgard RT Relay Module: • common horn relay • common trouble relay •...
Note that during the Warm-Up state of the Chillgard RT Refrigerant Monitor, all alarm relays in the Relay Module are cleared. Alarm Relay Configuration Latching The configuration of all alarm relays is controlled by the Chillgard RT Refrigerant Monitor. • If the Chillgard RT alarm relay (CAUTION, WARNING, ALARM) is configured as latching, all eight relays of the corresponding bank in the Relay Module are configured for latching operation.
Trouble Relay While a trouble condition in the Chillgard RT Refrigerant Monitor activates the trouble relay in the Chillgard RT Relay Module, a trouble condition in the Relay Module does not activate the trouble relay in the controlling Chillgard RT Refrigerant Monitor.
Horn Operation Configuration The configuration of the horn relay is controlled by the Chillgard RT Refrigerant Monitor. • The Horn Relay in the Relay Module is configured by the “Caution sets Audio Alarm,” “Warning sets Audio Alarm,” “Alarm sets Audio Alarm”...
Loss of communication Check wiring between Chillgard RT flashes Refrigerant Monitor and Relay Module Check power wiring or power supply Check wiring in the Chillgard RT Relay Module Power LED Power fault Check wiring, power supply and fuse not ON...
232 Output Introduction Your Chillgard RT Gas Monitor is ordered and built for the particular types of gas in use. The monitor full scale range may differ from the standard 0 to 1000 ppm range. Check the calibration label on the reverse side of the front cover (FIGURE C-1) to determine your instrument’s full scale range.
The instrument is ready to monitor; however, it is necessary to configure the unit to your specific requirements. Your Chillgard RT Gas Monitor may contain an internal Multi-Point Sequencer, enabling your unit to sample from up to eight individual sampling locations. Your unit also contains several other features described in this appendix: •...
Data Logged Value The data logged value is the next message. • The first line indicates the date and time. • The second line shows the: • point • gas type and • logged concentration. • The message is generated for every point and sent every 15 minutes or hourly, depending on the setup selection.
Table C-4. Alarm Logging RS-232 Output The RS-232 output broadcasts certain information about the Chillgard RT Gas Monitor (TABLE C-3). This output conforms to the specification for RS-232 signal levels and is capable of driving its signal up to 200 feet (61 m) when using low capacitance RS-232 cable. See FIGURE C-1 for RS-232 connector location.
Figure C-2. Typical RS-232 Wiring When connecting RS-232 cable to the instrument: • The strain relief cover or shell on the 25-pin sub "D" connector must be removed. It is not possible to use a pre-wired or molded cable because its shell or strain relief is not removable. The connector’s shell or strain relief does not fit within the instrument.
Remote Gas Sampling Channel Identification Chillgard RT Monitors have two analog outputs: • 4-20 mA and • 0-10 Volts. The voltage output can be programmed to correspond to the channel presently being sampled. The front panel screen indicates: 1V/PT IF OUT = 1-10V (1 volt per point if output equals one to 10 volts).
TABLE D-1. Introduction The Chillgard RT Refrigerant Monitor is available with the capability to detect up to six different refrigerant gases. This capability is used when multiple chillers using different refrigerants are located together or in separate machinery rooms.
After your unit is installed, open the front door and locate the Calibration Label (FIGURES D-1 and D-2). Also see the Manufacturing Date Label (FIGURES D-1 and D-3). If the Chillgard RT Refrigerant Monitor is built with multiple refrigerant gas capability, the calibration label shows what gases the unit can detect.
The only variation of the calibration procedure for the multi-gas Chillgard RT Refrigerant Monitor is to apply span gas to only one of the sample points. The particular sampling point chosen must be configured to monitor that particular gas.
Page 119
The screen in this section allows the user to define the refrigerant gas that is expected at each sampling point. Also, they determine which refrigerant gas is used for calibration by selecting the sampling point configured for that particular gas. Following are sample screens that differ from the Section 3, "Display Screens":...
Need help?
Do you have a question about the Chillgard RT and is the answer not in the manual?
Questions and answers