System:serror:list[:Next]; System:sutime; System:tlevels; System:transaction:begin - R&S NRP Series User Manual

Average power sensors
Hide thumbs Also See for NRP Series:
Table of Contents

Advertisement

®
R&S
NRPxxA(N)
Example:
Usage:

SYSTem:SERRor:LIST[:NEXT]?

Queries the list of static error changes for the oldest entry and removes it from the
queue.
Returns an error number and a short description of the error.
Example:
Usage:
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode, INITiate:CONTinuousON.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time>

SYSTem:TLEVels?

Queries the possible power test levels of the power sensor.
Usage:

SYSTem:TRANsaction:BEGin

Starts a series of settings.
Usage:
User Manual 1177.6017.02 ─ 08
SYST:SERR:LIST:ALL?
Response: 0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification".
Query only
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Query only
Range:
0.0 to 10.0
*RST:
10e-3
Default unit: Seconds
Query only
Event
Remote control commands
Configuring the system
116

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp6aNrp6anNrp18aNrp18an

Table of Contents