11.0 Notes on EMC
Immunity Test
Voltage Dips / Dropout
IEC 61000-4-11
NOTE
U
11.3 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS S 201 Thorax is intended for use in the electromagnetic environment specified below. The customer or
user of the ATMOS S 201 Thorax should ensure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
28
IEC 60601-
Test Level
< 5 % U
T
(> 95 % Dip of the U
for 0.5 Cycle
40 % U
T
(60% Dip of the U
for 5 Cycles
70% U
T
(30 % Dip of the U
for 25 Cycles
< 5 % U
T
(>95 % Dip of the U
for 5 s
is the mains alternating current prior to application of the test levels.
T
IEC 60601-
Test Level
V
= 3 V
1
eff
150 kHz to 80 MHz
E
= 3 V/m
1
80 MHz to 2,5 GHz
Compliance Level
< 5 % U
T
)
(> 95 % Dip of the U
T
for 0.5 Cycle
40 % U
T
)
(60% Dip of the U
T
for 5 Cycles
70% U
T
)
(30 % Dip of the U
T
for 25 Cycles
< 5 % U
T
)
(>95 % Dip of the U
T
for 5 s
Compliance Level
3 V
3 V/m
Electromagnetic Environ-
ment - Guidance
Mains power quality should be that
)
of a typical commercial or hospital
T
environment. If the user of the
ATMOS S 201 Thorax demands
continued function even in case of
)
interruptions of the energy supply,
T
it is recommended to supply the
ATMOS S 201 Thorax from an
uninterruptible current supply or a
)
battery.
T
)
T
Electromagnetic Environment -
Guidance
Portable and mobile communications
equipment should be separated from the
ATMOS S 201 Thorax incl. the cables by
no less than the distances calculated/listed
below.
Recommended distances:
d = [ 3,5 / V
] √P
1
] √P
d = [ 3,5 / E
1
80 MHz to 800 MHz
d = [ 7,0 / E
] √P
1
800 MHz to 2500 MHz
where „P" is the max. power in watts (W)
and D is the recommended separation
distance in meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compli-
ance level (b).
Interference may occur in the vicinity of
equipment containing following symbol.