3.3

Functional test

Mounting of the ATMOS S 201 Thorax as described
under Fig. 2.
Before use the system must always be tested regarding
any leakage.
Slip on the test hose set to the device and connect the
patient side of the hose to the sealing cap (blue plastic
part) at the secretion container. Use gloves!
Start the devices in pressing the key
is performed automatically by the device. Please press
Enter
to continue.
The permeability can be checked over a longer period
when the device has reached the Standard Screen (Fig.
8). Switch the pump on „ ► „ by briefly pressing button
when the target vacuum is achieved, switch
the pump off „ II " by again briefly pressing button 4.
The actual vacuum should not fall by more than 12 mbar
within10 secs when the hose end is closed.
In case the leakage test was successful Fig 4 appears
on the display. Please confirm with key
If the leakage test was not successful, please switch off
the device and check all the connections (hose, contai-
ner, etc.). Restart leakage test in starting up the device.
If need be re-establish permeability
(check hose connections).
Switch over to standard mode
is off „ II ",
3.4

SD card recording

After the leakage test the readout as shown in figure 5 or
figure 6 appears automatically on the display.
The readout shown in figure 5 indicates that no SD card
is in the device or a new card is being used. Please
insert the SD card and select "read in SD card" again
or select "therapy start with new recording" using the ▲
and ▼ buttons.
In case therapy should not be recorded, please select
"therapy start without recording".
When the SD card is already inserted prior to start up,
the readout as shown in figure 6 appears on the display.
It is possible to select either "continue previous recor-
ding" or "start new recording" with the ▲ and ▼ buttons.
.
A leakage test
(Fig 3).