4.0
Operation
Fig 12
Fig 13
4.6

Connect hose set

Measuring channel
Suction channel
Sealing cap for closing the secretion channel
Measuring channel:
Attach the luer connection with integrated filter to the measu-
ring channel and turn it to the right.
Suction channel:
!
Attach the luer connection to the measuring channel and turn
it to the right.
The vacuum is measured directly with the measuring channel
at the drain. A bacterial filter prevents the measuring channel
from contamination with bacteria and the penetration into
the device. The secretion is sucked off through the suction
channel.
A double-lumen hose system is indispensable! Both hoses
must be connected. The hose set is a single-use product for
use with one patient only.
!
When using other hose systems this may lead to damage of
the device.
4.7
Universal fastening (optional)
The fastening can be mounted to infustion tripods, wheelchairs,
to the bed or to the standard rail. Arrest the adjusting lever ,
hereto please pull out the pin  and rotate the adjusting lever
 up to the required position (in 90° steps) until the pin 
engages automatically. Fix the adjusting lever .
Please take care that the screw is screwed in prior to put the
!
ATMOS S 201 Thorax onto the support!
Put the ATMOS S 201 Thorax onto the support.
Slightly pull out the upper button  and rotate it by 90° until the
two connectors  are at the level of the holes  and then
release the button.