For Your Safety - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

2.0

For your safety

For your safety
The ATMOS S 201 Thorax is designed in accordance with
IEC 60601-1/EN 60601-1. It is a device which meets the
VDE protection class II. It may be connected to a properly
installed wall power point socket.
Before commissioning the device, drainage container,
mains cables, accessories, connecting leads and hoses
should be checked for damage.
Damaged leads and hoses must be replaced immediately.
The function of the device should be checked prior to
use.
The ATMOS S 201 Thorax may only be used by trained
specialist staff.
The ATMOS S 201 Thorax is not designed for operation
within areas where there is a risk of explosion (M and
G) and in oxygen rich environments. Areas that are at risk
from explosions can arise due to the use of combustible
anaesthetics, skin cleaning and skin disinfecting materi-
als.
No liquid may enter the device. If liquid has entered the
device it must be checked by the customer service before
it is used again.
After a long journey at temperatures under the freezing
point the device must be allowed to stand for up to six hours
at room temperature before it is used for the first time. If
the device is not acclimatised, it must not be operated,
since the piston of the unit might be damaged if it runs in
this condition.
Only trained staff may remove the container from the
device during therapy.
Prior to use the perfect function of optical and acoustic
alarm must be checked by the user.
Prior to use the nurse call function (menu item) must be
tested.
For hygienic reason we recommend to exchanged the
hose always together with the container.
Uncoupling of the device may only be performed if the
maximum target value of 15 mbar is reached.
Disinfectant should not enter the device. If disinfectant
has penetrated the device, then it must be dried tho-
roughly and subsequently an efficiency control must be
conducted. It must be both checked, whether the target
vacuum is reached when the system is closed and
whether a flow >10 l/min is reached after a while when
the system is open. If not, the device must undergo a
service check prior to being used again.
8
!
Dispose of the packing material in a proper manner.
Before the device is connected a check must be made to
see that the mains voltage and mains frequency given on
the device agree with the values of the supply system.
Only regular and undamaged mains connection and
extension cables must be used.
The suction hose must not come directly into contact with
the suction place but only through a sterile drain.
To disconnect the device from the mains first pull out
the plug from the wall socket. Only then disconnect the
connection cable from the device. Never touch the plug or
cable with wet hands.
The environmental conditions given in the Technical
Specifications (section 9.0) must be observed.
The control panel should always be clearly in view for the
operators and be easy to reach.
The ATMOS S 201 Thorax meets the resistance to jamming
of the standard IEC 60601-1-2/EN 60601-1-2 „Electro-
magnetic compatibility - Medical electrical devices".
No guarantee claims can be accepted for damage which
is produced by using third party accessories or third party
consumable materials.
ATMOS is not responsible for damage to people or goods
if
• non-original ATMOS parts are used,
• the information in these operating instructions are
disregarded,
• assembly, new installations, modifications, extensions
and repairs are done by people who are not authorised
by ATMOS.
These operating instructions correspond to the design of
the device and the situation of the safety standards on
which they are based when they were printed. The circuits,
processes, names, software programmes and devices are
all covered by patents.
Please only use ATMOS containers with integrated bacteria
filter and ATMOS hoses with integrated bacteria filter in the
measuring channel.
The container may not be used without device.
This product is not re-sterilisable. Repeated reuse of
components which are marked with a
In case of repeated reuse these components lose their
function and there is a high infection risk.
Attention loss of data
Removal of the SD card during operation of the thoracic
drainage device means that the currently written file is
deleted! The actual therapy recording will be lost!
2
is forbidden.

Advertisement

Table of Contents
loading

Table of Contents