Removal
3.3.5

Attachment of the universal bracket (Accessories)

Depending on the negative pressure chosen, location and course of the hose, a secretion col-
umn in the hose can cause the negative pressure in the wound to not correspond to the selected
value. Therefore, always place the vacuum unit at the same level or lower than the patient.
Turning the holder clamp
The stand holder may be attached – horizontally and vertically - to plates (e.g. table boards),
pipes and stands with a diameter up to 40 mm.
Ensure that the fixing pin is correctly clicked into place before attaching the vacuum unit on
top.
16
• Press the release button of the hose connector
laterally and keep it pressed.
• Now pull the two ends apart.
• Prior to disposal break off the sealing plugs on the
connector and plug it into the secretion hose on the
device. This will ensure that the secretion cannot
leak from the canister.
• The hose must not be clamped. Please do not use
any clamps.
• Pull the fixing pin from the fixing at the bottom side of the
stand holder.
• Turn the holder clamp until the fixing pin clicks into place
at the next fixing.