Troubleshooting - Atmos S 042 NPWT Operating Instructions Manual

Table of Contents

Advertisement

7.0

Troubleshooting

Description
Device cannot be
switched on.
Battery does not re-
charge and charging
symbol does not light
up despite proper
connection to the
power cable.
„Battery discharged"
„Hose blocked"
„Automatic switching
off"
„Secretion canister
full"
„Leakage"
"Battery service life
exceeded"
Device is hot
Possible causes
Battery is completely empty.
Power cable defective or not
connected properly.
Power supply or internal
battery is
defective
Battery almost empty.
Kinks in the hosing system.
Blockages at the connection
points.
Battery empty
Therapy not started
Wound dressing has a leakage Check wound dressing for any leakage
Connection between secretion
canister and vacuum unit has
a leakage.
Pump does not work
Liquid in the device
Service life of the battery is
reached
Remedy
Connect the power cable to the device
to recharge the battery, note that the
progress can be checked in the upper
left hand corner of the display.
Check the power cable.
Contact the ATMOS service or a certi-
fied service partner.
Connect device to the supply network.
The battery is charged and the state of
charge is indicated in the display.
Remove the kink.
Check connections resp. exchange the
secretion canister.
Connect device to the supply network.
The battery is charged and the state of
charge is indicated in the display.
Switch on pump.
Exchange secretion canister.
if need be exchange the dressing.
Check connection between vacuum
unit and canister, if need be replace
secretion canister.
Contact the ATMOS service or a certi-
fied service partner.
Switch on the device without the secre-
tion canister and start the therapy.
Suppress the error message and run
the device until the next error mes-
sage appears.
Subsequently conduct an efficiency
control (device with canister).
a) device reaches the target vacuum
and can be used again by the patient.
b) device does not reach the target
vacuum.
Contact the ATMOS service or a certi-
fied service partner.
Contact the ATMOS service or a certi-
fied service partner.
Please ensure sufficient air ventilation.
29