4.1.4

Switching the intermittent mode on and off

With your negative pressure unit, you have the option of selecting between continuous and
intermittent mode. In contrast to the continuous mode, which operates with a negative pressure
that remains the same, the intermittent mode allows therapy with changing pressure intervals.
In the intermittent mode the difference between the set vacuum values must be at least -20
mmHg. A smaller difference is prevented by the negative pressure unit.
How to adjust the preset values
• Touch the value you would like to change with a finger. The respective field is marked in dark
blue.
• Now touch the
or
• To save the newly set values touch the
The values are only saved if the intermittent menu is closed with the
wish to accept the values, you have set, wait until the screen automatically returns to the main
menu. But do not touch the screen for this.
4.1.5

Key lock

The ATMOS
S 042 NPWT is equipped with an automatic key lock.
®
1. Automatic activation of the key lock
If no adjustments are made to the settings on the screen over a period of one minute then the
key lock is automatically activated. This prevents accidental operation of the device. The sym-
bol
appears in the display.
When the key lock is activated the display turns black after 5 minutes.
Switching the intermittent mode on
To switch on the intermittent mode press the
key appears on the screen. This key indicates the
120
3
20
3
mmHg
Min.
set values of the intermittent mode and is framed in green.
Change values
To adapt the preset values touch the
You reach the intermittent menu.
key to set the desired value.
key.
2. Manually activate the key lock
The key lock can be manually activated after all the therapy values
are set and the therapy is started. When the
can activate the key lock.
Press the
symbol to activate the key lock. When the key lock is
activated the display turns black after 5 minutes.