Table A- 3 Input/Output Wiring (Tb1) - Siemens SITRANS F Series Operating Instructions Manual

Ultrasonic flowmeter
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Table A- 3
Pin#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
FST020 IP65 NEMA 4X
Operating Instructions, 02/2012, A5E03086487-05
Input/Output Wiring (TB1)
Signal
Io1+
Isolated Loop Supply
Io1-
Isolated Loop Return
GND
Ground
C
C
NO
Normally Open
NC
Normally Closed
GND
Ground
NO TOT +
DIGITAL INPUT +
NO TOT -
DIGITAL INPUT -
GND
Ground
CLR TOT +
DIGITAL INPUT +
CLR TOT -
DIGITAL INPUT -
GND
Ground
PULSE OUTPUT +
Isolated Transistor
PULSE OUTPUT -
Isolated Transistor
GND
Ground
A.3 I/O Connections and Wiring
Function
Spannable 4-20mA output (Loop
Powered) This output also provides a
fault indication by dropping to 2mA if
assigned to flow rate and under fault
conditions.
Ground
Relay Output
Relay Output
Relay Output
Ground
Stops Totalizer from incrementing.
Ground
Clears Totalizer
Ground
Menu selection: PGEN, POS TOTAL,
NEG TOTAL
Ground
Appendix
Description
123

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fst020 ip65 nema 4x 7me3570

Table of Contents