Download Print this page

Keysight Technologies B1500A Self-Paced Training Manual page 90

Semiconductor device analyzer
Hide thumbs Also See for B1500A:

Advertisement

Module 2
Basic Measurement
The 16058A has 4 non-Kelvin SMU inputs and 3 coaxial inputs for the other instrument connection.
By adding the 16493A Adapter Cable, you get access to 5 more coaxial connections. The adapter
was designed to bring out the fixture interlock connection and 4 additional connections, such as the
4142/4155/4156's VSU/VMU.
The 16493A cable converts the 16058A fixture interlock connection to a single coaxial connector.
This mates directly with the 16435A Interlock Cable Adapter mentioned on the previous page.
Note
You will notice that the socket modules for Keysight Technologies entire DC parametric product line
are interchangeable. So socket modules that came with the 4142, 4145, 4155, or 4156 will work with
the 16058A, 16088A/B, and 16442A/B fixtures.
2-62

Advertisement

loading