Module 5
SPGU Control and Applications
To apply the ALWG output, define the ALWG output sequence, specify the Operation, and click the
Single button.
The following operations are available.
PULSE COUNT: SPGU outputs the specified number of sequences. Specify a number within the
range of 1 to 1000000.
DURATION: SPGU continues the output until the specified time elapses. Specify a time within the
range of 1 us to 31556926 s (1 year).
FREE RUN: SPGU continues the output until the measurement is completed or stopped. This
operation is not available for the SPGU Control classic test.
To define the ALWG output sequence,
1. Open the SPGU ALWG Setup window by clicking the Pulse/ALWG button on the SPGU Control
classic test screen or the SPGU ALWG Setup button on the Measurement Setup screen of the other
classic test.
2. Open the Define ALWG Waveform window by clicking the button of the ALWG Source on the
SPGU ALWG Setup window.
The ALWG output sequence can be defined by specifying the waveform pattern and the repeat count
on the Define ALWG Waveform window. To create a new waveform pattern, use the Pattern Editor.
See the next page.
5-34