Underfloor Installation Of The Pc Connections - Dentsply Sirona TENEO Installation Requirements

Table of Contents

Advertisement

Dentsply Sirona
Installation Requirements Teneo
61 93 739 D3509
D3509.021.01.18.02 04.2018
3.6

Underfloor installation of the PC connections

Depending on the prevailing local conditions, the existing cable set can
be installed in the cable duct of an underfloor installation by an installer.
Cable channel No. 9 is used for this purpose, see Installation template
[ → 17].
IMPORTANT
The cable set for the PC connection must not be routed in the same
cable channel as any existing radiation cables for a HELIODENT Plus
unit model!
Cable sets for PC connection with HDMI and USB cable for camera
SiroCam
● 5 m: REF 64 46 111
● 10 m: REF 63 29 655
NOTE
Electric lines are susceptible to breakages.
Any kinks or twists in the cables could damage their wires. You must
then replace such cables.
➢ Ensure that electrical lines do not become kinked or twisted.
Running cables to the PC
Lines USB repeater, L339 (Ethernet), L406 (HDMI) and protective ground
wire. For PCs without a HDMI output, the Audio line is also required.
To prevent transmission interference, ensure that the cables are not
crossed.
A cable duct is laid from the treatment center to the location of the PC.
Free length A of cables at the treatment center end:
Length A = 600 mm
3 On-site installation
3.6 Underfloor installation of the PC connections
A
C
21