4 Dimensions, technical data
4.4 Technical data
34
Foot control wireless interface
The wireless foot control can be equipped with various wireless modules.
A label with the name of the wireless module used is on the underside of
the foot control.
Wireless module until April 2017:
Model
designation:
Frequency:
Transmitting
power:
Modulation type: MDMA
Range:
Approval:
Wireless module starting April 2017
Model
designation:
Frequency:
Transmitting
power:
Modulation type: GFSK
Range:
Approval:
IMPORTANT
Minimum requirements for the PC
See document "Installation instructions and system requirements for PC
configuration," (REF 61 94 075) Sivision Digital.
4.4.1

Operating, transport and storage conditions

Operating mode:
Installation Requirements Teneo
nanoLOC AVR
2.4 GHz – 2.4835 GHz (ISM band)
< 2 mW (short-range device)
approx. 10 m
See "Standards/Approvals" [ → 36].
BlueMod+S
2.4 GHz – 2.480 GHz
< 2 mW (short-range device)
approx. 10 m
See "Standards/Approvals" [ → 36].
Continuous operation with intermittent
loading corresponding to the dental
mode of working.
Drive motors for patient chair and
backrest: max. 25s on and 400s off
Other motors: max. 15s on and 3min off
Permanently connected unit. Operation
is not permitted in mobile vehicles.