Dentsply Sirona - Dentsply Sirona TENEO Installation Requirements

Table of Contents

Advertisement

3 On-site installation
3.4 Underfloor installation of supply lines
18
1
Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
2
Compressed air supply
Pipe 10x1 mm, corner valve outlet 3/8"
3
Suction line
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter approx. 36.5 mm)
4
Water drainage
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter approx. 36.5 mm)
5
Installation pipe (power supplies)
inner diameter approx. 40 mm)
6
Control cable to relays for the
suction machine (
%
call cable (
),
Special function (#)
2
3 x 1.5 mm
(AWG16) (quality as in the power cable)
7
Power supply
2
3x1.5 mm
(AWG16)
Circuit breaker: for 230 V AC: 16 A medium-blow
for 100 - 115 V AC: 20 A medium-blow
Recommendation: Type B automatic circuit breaker
8
Option: HELIODENT Plus radiation cable (10 m)
(free end 800 ±50 mm)
CAUTION
The HELIODENT Plus radiation cable cannot be laid
together with USB or HDMI cables. If required, allow for a
separate installation pipe or above-floor installation, see
Above-floor installation of supply lines.
9
Installation pipe (IT)
inner diameter min. 50 mm
(or corresponding flat conduit) for the PC connection and
CEREC AI connecting cable.