Safety information

2.1

Installation by qualified personnel

The installation of the supply connections must be carried out by qualified
technicians only.
Electrical and water supply installations; reference for INST suction machines
WARNING
Professional installation
Comply with the national regulations for electrical installation
(e.g. IEC 60364-1, VDE 0100-100, IEC 60364-7-710, VDE 0100-710,
National Electrical Code).
Comply with the national regulations for water supply installations
(e.g. EN 1717) and sewage installations (e.g. EN 12056-1).
2.2

Radiotelephones

Mobile RF communications equipment can affect electro-medical
equipment. Therefore, the use of mobile wireless phones in medical office
or hospital environments must be prohibited.
2.3

Modifications and extensions of the system

Modifications to this device which might affect the safety of the system
owner, patients or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with
original Dentsply Sirona accessories or third-party accessories approved
by Dentsply Sirona. The user assumes the risk of using non-approved
accessories.
If any devices not approved by Dentsply Sirona are connected, they must
comply with the applicable standards, e.g.:
● IEC 60950-1 or IEC 62368-1 for information technology equipment,
(e.g. PC, Ethernet switch, Ethernet hub, Ethernet converter, USB
repeater, Ethernet router, USB switch, HDMI repeater/amplifier,
audio sources, display port repeater/amplifier. etc. and their power
adaptors), and
● IEC 60601-1 for medical electrical equipment.
The treatment center monitor must comply with the requirements of
IEC 60950-1 or IEC 62368-1.
The loudspeaker port of the monitor may be connected only to a device
that complies with IEC 60950-1, IEC 62368-1 (e.g., PC) or IEC 60601-
1.Under no circumstances should it be connected e.g., to a stereo
system.
If a system is created during the installation process, the requirements of
IEC 60601-1, 3rd edition, must be complied with. The person assembling
the system assumes responsibility for ensuring conformity, e.g., acc. to
Directive 93/42/EEC.
2 Safety information
2.1 Installation by qualified personnel
7