Underfloor installation of the PC connections

3.7
Depending on the prevailing local conditions, the existing cable set can
be installed in the cable duct of an underfloor installation by an installer.
Cable channel No. 9 is used for this purpose, see Installation template 
[→ 22].
IMPORTANT
The cable set for the PC connection must not be routed in the same
cable channel as any existing radiation cables for a HELIODENT Plus
unit model!
Cable sets for PC connection with HDMI and USB cable for camera
SiroCam are:
● 5 m: REF 64 46 111
● 10 m: REF 63 29 655
NOTE
Electric lines are susceptible to breakages.
Any kinks or twists in the cables could damage their wires. You must
then replace such cables.
➢ Ensure that electrical lines do not become kinked or twisted.
Running cables to the PC
5 m cables: L530 (USB), L553 (Ethernet), L552 (HDMI) and protective
ground wire.
10 m cables: USB repeater cable, L339 (Ethernet), L406 (HDMI) and
protective ground wire. For PCs with no HDMI output, the audio cable is
required, which is part of the cable set.
To prevent transmission interference, ensure that the cables are not
crossed.
ü A cable duct is laid from the treatment center to the location of the
PC.
ü Free length A of cables at the treatment center end: Length A = 600
mm
3 On-site installation
3.7 Underfloor installation of the PC connections