Sirona Windows Diagnostics - Dentsply Sirona CEREC AC Operating Instructions Manual

With cerec omnicam
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC AC
63 81 136 D3492
D3492.201.06.35.02    09.2019
10.2.3

Sirona Windows Diagnostics

General
The "Sirona Windows Diagnostics" test loads the PC-specific
components simultaneously over a long period of time. This period of
time is determined by the user.
As a result of this load
● the temperature in the PC is significantly increased compared with
normal operation.
A temperature malfunction or any existing faults are thereby
detected.
● Thanks to the temporal, unlimited test phase, sporadically occurring
faults are more likely to be detected.
Tip: Only conduct the test if at least one of the "Customer
Diagnostics" or "Technician Diagnostics" tests has been
completed with a "pass" and without errors.
Performing the test
Tip: Allow the "Sirona Windows Diagnostics" test to run for at
least 1 hour. For optimum results, allow the test to run overnight.
ü You have started the "Sirona Windows Diagnostics" test
and Windows starts in the test environment. The "Sirona
Windows Diagnostics" test starts automatically.
➢ Check the result in the "Windows Stress Test" window.
Tip: You may have to rearrange the windows to be able to see the
"Windows Stress Test" window.
10 PC Diagnostic Tool
10.2 Test procedure
81