Datex-Ohmeda F-LM1 Technical Reference Manual page 133

Light monitor
Table of Contents

Advertisement

Batt. last cap
Batt. last cap indicates last measured full capacity of the battery. The capacity is calculated from
a battery that is discharged from full to empty.
Batt. used cap indicates capacity discharged from the battery. Remaining capacity: last cap -
Batt. used cap
used cap.
Average voltage Average voltage indicates average of the last two load voltage measurements in battery use mode.
Batt1 cycles indicates number of charging times of battery 1.
Batt1 cycles
Batt2 cycles indicates number of charging times of battery 2.
Batt2 cycles
Service menu