Put the flex cable through the EMI filter.
NOTE: Work very carefully with the flex cable which breaks easily
Figure 7
Pull out the locking piece before pushing the flex cable into the connector.
After pushing the flex cable into the connector, push the lock piece back.
The flex cable does not make a contact with the connector, if the lock piece is loose
Figure 8
6
Document No. 896551-2
Flex cable and the EMI filter