Datex-Ohmeda F-LM1 Technical Reference Manual page 204

Light monitor
Table of Contents

Advertisement

Datex-Ohmeda Light Monitor
Notes
TEMP measurement
14. Communication and
memories
16. Calibration check
Notes
For F-LMP1 with the InvBP measurement
18. Cable and transducer
detection
20. Test with patient
simulator
Notes
SpO
measurement
2
21. SpO
probe detection
2
Notes
NIBP measurement
23. Communication and
memories
25. Leak test
26. Calibration check
0 mmHg
100 mmHg
200 mmHg
260 mmHg
z.o. = zero offset at 0 mmHg pressure
27. Watchdog timer activation pressure
28. Watchdog timer
Adult
A - 2(3)
Document No. 8000060-1
OK
N.A.
Fail
Measured B1
15. Temperature probe
detection
17. Temp test function
19. Calibration
22. Test measurement
24. Pump and valves
Measured
Measured B2
Measured
OK
N.A.
Fail
S/N
S/N
S/N
Allowed range
< 6 mmHg/min
Allowed range
± 9 mmHg
100 + z.o. ±2 mmHg
200 + z.o. ±3 mmHg
260 + z.o. ±4 mmHg
Allowed range
11-15 mmHg
120-140 s