Datex-Ohmeda F-LM1 Technical Reference Manual page 37

Light monitor
Table of Contents

Advertisement

8.
Check that the message "No CO2 sensor" is displayed when there is no CO
connected. Connect a CO
displayed.
9.
Wait until the "warming" message disappears and the CO
Breathe at least 5 times to the CO
are reasonable and the Respiratory value is calculated.
Fill in all necessary documents.
sensor to the monitor and check that the message "warming" is
2
sensor adapter and make sure that the measured values
2
Functional check
sensor
2
value appears on the screen.
2
Document No. M1051339
29