Datex-Ohmeda S/5 Light Monitor
Bad c-s by mod is a cumulative number that indicates how many communication errors the
module has detected.
The Light Monitor starts counting these items at power up and resets to zero at power off. The
nonzero values do not indicate a failure, but the continuous counting (more than 50 per second)
indicates either a serial communication failure, or a board not in place. Also LM-NIBP/LM-STP can
cause communication errors that cause these numbers rise.
RAM indicates the state of the external RAM memory.
ROM indicates whether the checksum at the EPROM is in accordance with the one the software has
calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either OK, Fail or ? (module not in place or a communication error).
60