Prepare to insert the new Lamp Units.
• Open the package and take the new Lamp Units out.
• Check the new Lamp Units for dust and foreign matters.
8)
Stand the TFT-LCD module and insert the new Lamp Unit.
The light guide and reflector sheet should be inside of the Lamp Unit.
Please pay attention to insert direction.
        
                                                    
9)
After inserting, please check the Lamp Unit is fastened with the latch.
T-51638D084J-FW-A-AB (AB) No. 2004-0162
Picture 7
The white cable should be on
the side of LCD Panel Surface.
Push down the latch of the plastic
frame (black) and insert the new
Lamp Unit.
The light guide and reflector sheet
should be inside of the Lamp Unit
keeping the white cable straight.
Please check the Lamp Unit
is fastened with the latch.
Picture 8