Datex-Ohmeda S/5 Light Monitor
1.3.4.2 Connector for network location ID plug, X6
The coding element interface is used for bedside address coding for the ethernet. The element is
plugged in to the female D-connector of 9-pins.
Pin out of the D9 female connector (front view)
Coding element pin assignments
Pin
I/O
1
O
2
O
3
4
O
5
O
6
O
7
O
8
O
9
8
Document No. 8000060-1
Signal
IDCS1
IDCL
I
IDDI
IDDO
IDPE
P5VD
Vcc for the coding element
ECGOUT
RESERVED, NOT IN USE
NCALL
RESERVED, NOT IN USE
gnd
Description
Chip select
Clock
Data in
Data out
Protect enable
(non-isolated)
Ground
Signal levels
CMOS 5V
CMOS 5V
CMOS 5V
CMOS 5V
CMOS 5V
see: module voltages