Datex-Ohmeda F-LM1 Technical Reference Manual page 74

Light monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Light Monitor
Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
The S/5 Light Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the S/5 Light Monitor should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
1)
±1 kV for mainstream CO
C-2(4)
Document No. M1051339
IEC 60601 test
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A/m
measurement
2
Compliance level
±6 kV contact
±8 kV air
1)
±2 kV or ±1 kV
for
power supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment. In case of disturbance
it is recommended to use transient
filter on power input lines.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If user of the S/5 Light
Monitor requires continued operation
during power mains interruptions, it
is recommended that the S/5 Light
Monitor be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic field
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.