both screws on the patient panel are tightened properly
3.
General functional inspection
Connect the power cord to the Power Adapter and the Power Adapter cable to the Light
Monitor.
Check that the green LED turns on in the power adapter and the LED on the lower right-hand
corner of the monitor turns on or starts flashing.
Check that both Battery charge status LEDs turn on or start flashing.
Switch the monitor on. Check that the monitor starts up properly, i.e.
both alarm LEDs turn on and off
the fan starts running
the start-up sound is heard from the loudspeaker
the normal monitoring screen appears
no error messages appear on the screen *)
if the monitor contains a recorder, two lines of start-up information should be recorded
check that the time and date are correct, adjust if necessary
if the monitor contains an F-LBAT or N-LBB, check that the battery symbol is displayed:
*) NOTE: "Check network connectors" will appear if the N-LNET or N-LDNET option is installed.
Enter the service menu and select Dummy Press.
Press the function keys one by one. Check that each key generates a sound from the
loudspeaker and the corresponding text.
4.
Screen setup
Configure the monitor screen according to the monitor configuration so that all necessary
parameter information is displayed, for example as follows:
Menu -
Monitor Setup
Connect the patient simulator to the monitor and check that all needed parameter
information is shown on the screen.
NOTE: The RESP waveform or Digit Field has to be selected on the screen before the respiration
measurement can be turned on.
-
Waveform Fields - Field 1 - ECG1
- Field 2 – P1
- Field 3 – P2
- Field 4 - Resp
- Digit Fields - Lower Field 1 - NIBP
- Lower Field 2 – CO2
- Lower Field 3 – Pleth
- Lower Field 4 – Temp
Service check
33
Document No. M1051339