NIBP error code explanation
CODE
EXPLANATION
0
RAM failure
Memory failure. Change NIBP board.
1
ROM checksum error
Memory failure. Change NIBP board.
2
+12V failure
Check short circuits. Change NIBP board.
3
-12V failure
Check short circuits. Change NIBP board.
6
ADC error
ADC circuit failure. Change NIBP board.
7
Watchdog time too short
Change NIBP board.
8
Watchdog time too long
Change NIBP board.
9
Watchdog activated
Change NIBP board.
10
EEPROM checksum error
Memory failure. Change NIBP board.
11
Auto zero range exceeded
Calibrate NIBP.
12
Communication break
Temporal break down of communication from monitor detected. Automatic recovery.
14
Too early Auto Start (needs 25 seconds without pressure)
Troubleshooting
41