Water Supply
If the upper limit of the water supply temperature to the processor is exceeded, the developer
temperature may not be controlled correctly. A water chiller may be required.
a.
Processor Supply
(1) Temperature: 4° to 32°C (40° to 90°F)
(2) Pressure: 172.35 to 448.11 kPa (25 to 65 psi) Install regulator if required.
(3) Volume: Controlled within the processor to 5.7 L/min. (1.5 gal/min.)
(4) Filtration: A 50-micron filter is required.
b.
Water service must comply with local codes.
c.
Tempered water service is suggested for processor cleaning and for mixing chemicals manually.
d.
A molded adapter and washer are provided to adapt the processor garden hose fitting to the 1/2 in. N.P.T.
(male).
Drain
• Capacity: 15 L/min. (4 gal/min.)
• Connection: Open drain; avoid solid connection
Do not use brass or copper for the drain lines. Drain service must comply with local codes.
1-6
SECTION 4
Water and Drain Requirements
NOTE
NOTE
246628