F100 K I L L - Reboot - Symmetricom XLi User Manual

Time & frequency system
Table of Contents

Advertisement

S
S
S
S
S
S
S
S
S
where:
= ASCII character F
F
= unit function number
100
= space
<S>
= specify User Name command
J
<CR> = input line terminator
The XLi responds:
F100 J FACTORY MODE=1<CR><LF>
or
F100 J FACTORY MODE=0<CR><LF>
F100 K I L L – Reboot
Use function F100 K I L L to reboot the unit. Use F100 K I L L after upgrading the system firmware.
K I L L is a case-sensitive command. When entering this command, use all capital letters and put
spaces between each letter.
To reboot the unit, enter:
F100 K<S>I<S>L<S>L<CR>
For example:
F100 K I L L<CR>
XLi responds:
OK <CR><LF>
RESETING THE UNIT<CR><LF>
PLEASE WAIT...<CR><LF>
In a network port session, rebooting the XLi terminates the network port session; open a new network
port session when the XLi has finished rebooting. In a serial port session, the XLi displays text similar to
the following example when the XLi has finished rebooting and is ready to receive additional commands:
>SYSTEM POWER ON SELF TEST RESULTS:
SERIAL LOOPBACK TEST PASSED.
RAM TEST PASSED.
PROG CRC TEST PASSED
NETWORK INTERFACE 192-8001
ALL RIGHTS RESERVED
FLASH FILE SYSTEM MOUNTED.
SOURCE FILE /config/truetime.conf BYTES READ: 1210
FILE SYSTEM REV #
XLi Time & Frequency System
XLi-man, Issue 8, 6/17/2008, Rev. H
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
S
S
S
S
S
S
S
S
S
(c) 1998 - 2008 SYMMETRICOM
1.80
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
2
1
5
S
S
S
S
133

Advertisement

Table of Contents
loading

Table of Contents