Reference Accuracy - Emerson Rosemount 3051 Manual

Transmitter with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 3051:
Table of Contents

Advertisement

Rosemount Model 3051 Transmitter with F
DETAILED
PERFORMANCE
SPECIFICATIONS

Reference Accuracy

Ambient Temperature per
50 °F (28 °C)
A-2
TM
fieldbus Protocol
OUNDATION
Zero-based calibrations, reference conditions, silicone oil fill, and 316
SST isolating diaphragm
Model 3051T/CA Ranges 1–5:
For calibrated ranges less than 10:1, accuracy =
 % of Calibrated Range
URL
--------------------------------------------- -
0.0075
Calibrated Range
Model 3051CA Range 0:
For calibrated ranges less than 5:1, accuracy=
Model 3051CD Ranges 1–5 and Model 3051CG:
URL
0.025
0.01
--------------------------------------------- -
+
Calibrated Range
Model 3051 CD Ranges 1-5 and Model 3051 CG:
For calibrated ranges less than 10:1
(15:1 for Model 3051CD Range 1), accuracy:
URL
±
-----------------------------------------
0.025
+
0.005
Calibrated Range
Model 3051 CD Range 0
For calibrated ranges less than 2:1 to 30:1,
accuracy = 0.05% URL
Model 3051L and Model 3051H
For calibrated ranges less than 10:1, accuracy =
URL
±
-----------------------------------------
0.025
+
0.005
Calibrated Range
Model 3051CD/CG
±(0.0125% URL + 0.0625% calibrated range) from 1:1 to 10:1
±(0.025% URL + 0.125% calibrated range) from 10:1 to 100:1
Range 1: ±(0.1% URL + 0.25% calibrated range)
Range 0: ±(0.25% URL + 0.05% calibrated range)
Model 3051L
See the Rosemount Instrument Toolkit
Model 3051T and Model 3051CA
±(0.025% URL + 0.125% calibrated range) from 1:1 to 30:1
±(0.035% URL + 0.125% calibrated range) from 30:1 to 100:1
Range 0:±(0.1% URL + 0.25% calibrated range)
Range 5:±(0.1% URL + 0.15% calibrated range)
% of Calibrated Rang
% of Calibrated Range
% of Calibrated Range
®
or SOAP 2000
software.