2. The first parameter is called "BLK_TYPE_n". This is a drop down
menu and the selections available in the device will appear. Select
the block to be displayed, in the "BLK_TYPE_n" field.
3. Next, select "BLK_TAG_n," enter the exact name of the block to be
to displayed. This must be the same name as the one that is stored
in the device or shown in the NI-Configurator screen.
4. Then select "PARAM_INDEX_n." Use NI_Dialog for this step. The
index number of the parameter that will be displayed on the LCD
is needed for this step. Using NI-Dialog, open the "GetParamList"
of the block in which the parameter resides. Count down the list
until the desired parameter is reached. This is the number used
next. Open the "GetParamList" list for the LCD block and write to
parameter "PARAM_INDEX_n" the value obtained in the above
parameter list, this must be in hex.
5. The next parameter is called "CUSTOM_TAG_n." This is an
optional selection that identifies which parameter, block or device
you are viewing on the LCD. This can be any name up to five
characters long.
6. Select "UNITS_TYPE_n". Select "Custom" in this parameter if
bringing a value from outside the Model 3051 device. "Auto" has
only pressure units, which might or might not match the disired
selection.
7. Select "CUSTOM_UNITS_n." This parameter is limited to five
characters and is where the desired units are displayed if
"Custom" was selected above.