OUNDATION
Power Supply
External power supply required; transmitters operate on 9.0 to 32.0 V dc
transmitter terminal voltage.
Current Draw
17.5 mA with all configurations (including LCD meter option)
Overpressure Limits
Transmitters withstand the following limits
without damage:
Model 3051CD/CG
Range 0: 750 psi (52 bar)
Range 1: 2000 psig (137,9 bar)
Ranges 2–5: 3626 psig (250 bar)
Model 3051CA
Range 0: 60 psia (4,1 bar)
Range 1:120 psia (8,3 bar)
Range 2: 300 psia (20,7 bar)
Range 3: 1600 psia (110 bar)
Range 4: 6000 psia (414 bar)
Model 3051H
All Ranges: 3626 psig (250 bar)
Model 3051TG/TA
Range 1: 750 psi (52 bar)
Range 2: 1500 psi (103 bar)
Range 3: 1600 psi (110 bar)
Range 4: 6000 psi (414 bar)
Range 5: 15000 psi (1034 bar)
NOTE
For Model 3051L or Level Flange Option Codes FA, FB, FC, and FD, limit is 0 psia to the flange rating or
sensor rating, whichever is lower.
Table A-4. Model 3051L and Level Flange Rating Limits
Carbon
Steel
Standard
Class
Rating
ANSI/AS
150
285
ME
psig
ANSI/AS
300
740
ME
psig
ANSI/AS
600
1480
ME
psig
DIN
PN
40 bar
10-40
DIN
PN
16 bar
10/16
DIN
PN
40 bar
25/40
(1) At 100 °F (38 °C), rating decreases with
increasing temperature
Stainless
Steel
Rating
275
(1)
(1)
psig
720
(1)
(1)
psig
1440
(1)
(1)
psig
(2)
(2)
40 bar
(2)
(2)
16 bar
(2)
(2)
40 bar

Advertisement