Emerson Rosemount 3051 Manual page 98

Transmitter with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 3051:
Table of Contents

Advertisement

Rosemount Model 3051 Transmitter with F
Table C-1. Transducer Block Parameters (continued)
Parameter
SENSOR_RANGE
SENSOR_SN
SENSOR_TYPE
ST_REV
STRATEGY
TAG_DESC
TB_DETAILED_STATUS
TRANSDUCER_DIRECTORY
TRANSDUCER_TYPE
UPDATE_EVT
XD_ERROR
Block/Transducer Errors
C-4
OUNDATION
Index
Number
Description
21
The high and low range limit values, the engineering units code, and the number of
digits to the right of the decimal point for the sensor.
22
Serial number of the sensor.
20
Type of sensor connected with the transducer block.
Valid sensor types are the following:
117 = Capacitance
124 = Strain Gauge
01
The revision level of the static data associated with the function block.
03
The strategy field can be used to identify grouping of blocks.
02
The user description of the intended application of the block.
31
Indicates the state of the transmitter. The parameter contains specific codes relating
to the transducer block and the pressure sensor specifically.
09
Directory that specifies the number and starting indices of the transducers in the
transducer block.
10
Identifies the transducer that follows.
100 = Standard pressure with calibration
07
This alert is generated by any change to the static data.
11
Provides additional error codes related to transducer blocks.
The following conditions are reported in the BLOCK_ERR and
XD_ERROR parameters. Conditions in bold type are available.
Conditions in italics are inactive for the Transducer block and are given
here only for your reference.
Table C-2. BLOCK_ERR and XD_ERR Conditions
Condition
Number
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
TM
fieldbus Protocol
Condition Name and Description
Other
Block Configuration Error
Link Configuration Error
Simulate Active
Local Override
Device Fault State Set
Device Needs Maintenance Soon
Input failure/process variable has bad status
Output Failure
Memory Failure
Lost Static Data
Lost NV Data
Readback Check Failed
Device Needs Maintenance Now
Power Up: The device was just powered-up.
Out of Service: The actual mode is out of service.

Advertisement