Installation Requirements

Parameter
Operating temperature range
Storage temperature range
Relative humidity
Storage humidity range
Bench space
Static electricity
Power
Surface orientation

Startup Kit Contents

The part number for the 2410 detector startup kit is 700000135. The kit contains the following
parts:
Lower Level P/N#
716000115
WAT048918
71500241002
WAT025604
WAT051053
WAT051329
WAT027893
WAT087198
WAT024193
WAT163-14
WAT270868
WAT024211

Startup Procedure

Requirement
+15 °C to +32.2 °C (59 °F to 90 °F); avoid direct exposure
to sunlight and heating/cooling vents.
–40 °C to 70 °C (–104 °F to 158 °F)
20% to 80%, noncondensing
0% to 90%, noncondensing
At least 11.2 in. (28.4 cm) wide × 24.8 in. (63 cm) deep ×
8.2 in. (20.8 cm) high (includes 5 in. [12.7 cm] clearance
at rear)
< 8 kV contact
Grounded ac, 100/240 Vac, 50/60 Hz
Level (ensures proper drip tray function)
STARTUP KIT LIST
CABLE RECORDER
OPERATORS MANUAL
COMPRESSION FITTING 5PK
TUBE TFE .058X.080X120
TUBE SS .062X.040X60
TUBE SS .062X.009X28
CABLE IEEE 3.3 FEET
BAG 3X4 ZIP LOCK
FUSE 2.0A 250V 5X20 QTY 2
CONN PLUG 12 PIN QTY 2
BAG, AIR BUBBLE