Transport & Storage -4; Electric Safety Information -4 - Hanil FLETA 4 Operation Manual

Clinical centrifuge
Hide thumbs Also See for FLETA 4:
Table of Contents

Advertisement

1. General Considerations
1.2 Transport & Storage
• The device and the accessories
may only be stored in dry rooms
-Storage-
Ambient temperature 5℃~ 40℃
Maximum relative humidity 30%~85%
Air pressure 500~1060hpa
1.3 Safety label attached to a product
Insert tube
symmetrically.
1.4 Electric safety information
1. It is recommended that switchgear or circuit breakers and overcurrent protection devices
be installed near the equipment.
2.Use a power cord only provided with equipment.
3.Use sockets with a protective earth conductor and suitable power cord.
4.Do not use an extension cord.
5.Do not place anything on the power cable
6.Do not block vents.
7.If you have the following emergencies, shut off the power supply and unplug the
power cord from outlet and contact your place of purchase.
-Unusual noises or smell from the equipment.
-Damage or wear of a power cord.
-Breakdown of circuit breaker, fuse or safety device.
-If you spill liquid on the equipment.
-If the device has been damaged.
Firmly tighten
the rotor lid.
Watch your figers
when close the lid.
A/S : +82-2-3452-8966 / techsupport@ihanil.com
• Only lift and transport the
device with sufficient number of
helpers.
-Transport-
Ambient temperature -10℃~40℃
Maximum relative humidity 10%~90%
Air pressure 500~1060hpa
Mark indicating danger
and warning.
Mark indicating a place in
danger of electric shock.
Mark indicating earth grounding
4

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents