LOW SPEED CENTRIFUGE T05R - Product name :Centrifuge - Model name : T05R - Manufacturer : Hanil Scientific Inc. - Address : 16 Arayukro, Gimpo, #10136, korea Reasearch Use Only www.ihanil.com UM-T05R(E)(Rev.2), 20210406...
USER MANUAL T05R 1. Safety Warnings and Conditions 1-1. Safety label 1-2. Conditions of use, storage or transport Electric shock hazard caution sign Caution sign indicating danger and warning Operate with all Insert tubes Assure the rotor locked safely Watch out buckets mounted.
If it is connected with the wrong voltage, it may cause damage to the device and personal injury. 3. Use only the rotor and recommended parts and accessories provided by Hanil Scientific Inc. We are not responsible for any damage to the device or accidents resulting from the use of non-recommended parts and accessories.
USER MANUAL T05R 2. Product Composition and Information 2-1. Appearance Tool Box Power Socket & Fuse Drain Knob Display & Control Power Swich Rotor- Optional Manual Lid Open Hole 2-2. Components Components T-wrench User Manual AC Power Cord Grease(Lubricant) USER MANUAL ihanil.com...
3. Product installation 3-1. Unpacking 1. After purchasing the centrifuge, open the box and check the components. Centrifuge (T05R) / User Manual / AC power cord / T-wrench for rotor locking and emergency door opening. 3-2. Power connection 1. Connect the AC power cord to the power socket located on the back of the main body and connect the power plug to the outlet.
USER MANUAL T05R 3-3. Lid Opening & Closure ▶When lid is opened, the lamp will be ON ▶When lid is closed, the lamp will be OFF The lid lock system of this product is locked with only a light touch, so do not apply excessive force.
USER MANUAL T05R 3-4-2. Fixed Angle Rotor. 1. Insert the rotor into the motor shaft and tighten it completely using the locking tool. ▶ Rotor tightening: Rotating with a Rotor Locking Tool and rotating clockwise ▶ Rotor Separation: Rotating with a Rotor Locking Tool and rotating counterclockwise ▶...
Page 11
USER MANUAL T05R 1. Before loading smaple tubes, check for any moisture or foreign substance in the rotor hole or bucket. ▶ If there is any, remove it with soft dry cloth. 2. Tubes must be placed in the rotor with same amount of samples symmetricaly.
USER MANUAL T05R 4. Operation 4-1. Key Functions of Control Panel 5000 09:59 4 □ RPM / RCF For automatic conversion of RPM / RCF and to set the speed Control interval : 1 RPM / 1 xg □ AT SET SPEED Use to count the run time once the actual run speed reaches to the set speed value □...
USER MANUAL T05R 4-2. Setting the RPM Value The speed setting value is displayed as RPM and RCF, and is automatically calculated correspondingly 1. Touch the [RPM / RCF] button once. 5000 09:59 4 ▶Touching the button once / twice →RPM / RCF setting...
Page 14
USER MANUAL T05R AT SET SPEED MODE For exact time control, this instrument can be set with AT SET SPEED mode which counts the run time once the actual run speed reaches to the set speed value and stops when the deceleration begins.
USER MANUAL T05R 4-4. Setting Temperature and Fast Cool The temperature setting range can be set from -10ºC to 40ºC For temperature sensitive samples, fast cooling is supported, reaching room temperature or bellow in a short period of time. 1. Touch [TEMP] button...
USER MANUAL T05R 4-5. Acceleration / Deceleration Set the acceleration rate to 9 steps and the deceleration rate to 10 steps (Natural Depreciation: 0) to protect sensitive samples and clean separation. 1. Touch [ACC/DCC] button 5000 09:59 2. Touch [ENTER] button after inputting set value to input ACC ▶...
USER MANUAL T05R 4-6. Start/Stop This button is used to start or stop centrifugation. The Start/Stop button lamp illuminates during operation. 4-6-1. Start 1. Touch the [Start/Stop] button after completing the speed, time, and temperature settings. ▶ During operation, the lamp of the right...
USER MANUAL T05R 4-7. Pulse Mode The [Pulse] button is for when press & hold, to accelerate up to maximum speed and decelerate when the button is not pressed. 1. Set the desired acceleration and deceleration 5000 09:59 4 steps.
USER MANUAL T05R Program Recalling Program recalling can call saved programs between 00 and 99 1. Touch [PROG] button once. Call 5000 ▶ 'Call' is shown in the display window. 2. Touch [ENTER] and enter the program number to be called.
USER MANUAL T05R 4-10. Setting sound volume Adjust the sound of the end of the action from 0 to 10 (0: mute). 1. Touch [Sound] ▶Touch [sound] → Display shows SOUND LEVEL Sound LEVEL_03 2. Touch [Enter], input set value to adjust volume ▶Touch [Enter] →...
USER MANUAL T05R 4-12. Emergency Lid-Lock Release The following is the procedure to remove samples when the lid cannot be opened automatically due to no power to the main body 1. Make sure the inner rotor is completely stationary. 2. Remove the manual lid opening cap on the bottom left side and check the hole.
USER MANUAL T05R 6-2. Error Code If the problem persists after taking the following measures, please contact the service center. ihanil.com A/S Inquiry : 02-3452-8966 / techsupport@ihanil.com...