6. Troubleshooting
6.2 Error codes
Error
Error
code
message
E12
Hi voltage
error
E13
Low voltage
error
E15
RPM Sensor
error
E17
Communicati
on error
Cause
- High voltage
- Low Voltage
- RPM Sensor error
- Rotor identification error
- Occurs when Main and IO
board are disconnected
A/S : +82-2-3452-8966 / techsupport@ihanil.com
Recommended action
- Shut off the power supply, and then, turn on
the power switch again to check the device.
- Contact Hanil Scientific Inc.
- Check the id disc is on the bottom of the
rotor
- Contact Hanil Scientific Inc.
20