Electrical Requirements; Single Phase Requirement - Agilent Technologies 5100 Site Preparation Manual

Icp-oes
Hide thumbs Also See for 5100:
Table of Contents

Advertisement

Laboratory Facilities

Electrical Requirements

Table 5. System electrical specifications
System unit
Spectrometer
PC, Monitor, and Printer
G8481A Water chiller
G8489A Water chiller
SPS 4 Autosampler
SPS 3 Sample
Preparation System
ASX-500 Series
Autosampler
VGA 77
SVS 2+
IsoMist

Single phase requirement

The Agilent ICP-OES requires a single phase mains input voltage of
between 200-240 VAC (50-60Hz) and draws a maximum of 15 amps
RMS at 200 to 220 volts and 13 amps at 230-240 volts (2.9 kVA) with
a power factor of approximately 0.99. (Can be a cross phase
connection to achieve specified supply voltage).
26
Required supply voltage
200–240 VAC Single Phase
50 Hz–60 Hz
100-127 VAC
200-240 VAC
220–240 VAC, 50 Hz
120 VAC, 60 Hz
240 VAC, 50 Hz
208-230 VAC, 60 Hz
100–240 VAC, 47–63 Hz, 1.5 A
100–240 VAC, 50–60 Hz
85-264 VAC
100 or 120 VAC,
220 VAC (equivalent to 230 VAC), or
240 VAC (equivalent to 230
VAC)Frequency 49–61 Hz
100–240 VAC, 50–60 Hz
100–240 VAC, 50-60 Hz
Agilent 5100 and 5110 ICP-OES Site Preparation Guide
Maximum Power
Consumption
200-220 V, 15 A
230-240 V, 13 A
10 A
5 A
8.9 A
16 A
12.2 A
12.2 A
<1 A
<1 A
Nominal rating
2.9 kVA
1000 VA
1000 VA
2140 VA
1920 VA
2900 VA
2900 VA
24 VDC, 2.5 A
~220 VA
40 VA
24 VA
1.8 A
2 A