Site Requirements and Specifications

Site Requirements

A suitable environment is important to ensure optimal performance of the
detector.
Power Consideration
The detector power supply has wide ranging capabilities and accepts any line
voltage in the range mentioned in
no voltage selector in the rear of the detector. There are also no externally
accessible fuses, because automatic electronic fuses are implemented in the
power supply.
To disconnect the detector from line, unplug the power cord. The power supply still
WA R N I N G
uses some power, even if the power switch on the front panel is turned off.
Shock hazard or damage of your instrumentation can result, if the devices are
WA R N I N G
connected to a line voltage higher than specified.
Make sure to have easy access to the power cable of the instrument, in order to
C A U T I O N
disconnect the instrument from line.
Power Cords
Different power cords are offered as options with the detector. The female end
of all power cords is identical. It plugs into the power-input socket at the rear
of the detector. The male end of each power cord is different and designed to
match the wall socket of a particular country or region.
34
Table 2
on page 36. Consequently, there is